Chemiluminescence describes the emission of light that occurs as a result of unique chemical reactions. In a chemiluminescence reaction, energy is released in the form of photons when electronically excited molecules, produced as a result of the reaction, relax to a stable ground state. The high signal/noise ratio that is achieved in these reactions makes chemiluminescence one of the most sensitive detection techniques used in medical diagnostics today.

 

Invitron’s Molecular Light Technology uses acridinium ester (AE) chemiluminescent compounds that can be linked directly to binding reagents such as antibodies or oligonucleotide probes. These labelled reagents can be individually designed to target an infinite range of substances of biomedical interest which are then quantified by the intensity of the light generated in the chemiluminescent reaction.

 

Our AE labelling compounds have enabled a new generation of advanced laboratory diagnostics, delivering higher diagnostic sensitivity and accuracy than previous techniques and delivering test results in a fraction of the time. For this reason AE technology has become the end-point-detection system of choice for leading clinical diagnostic analysers. Today, the technology still remains unsurpassed in terms of its sensitivity, precision and wide dynamic range. Invitron’s next-generation POC technology translates the functionality of an automated chemiluminescence analyser into a miniaturised format, bringing a new dimension to the range of applications for chemiluminescence assays.

Molecular Light Technology has received several accolades including a Queen’s Anniversary Prize and the US National Medal of Technology & Innovation. During the ‘GREAT for Imagination’ campaign, launched by the UK government to mark the 400th anniversary of the first British patent (GB1), the AE invention (GB2112779) was identified as one of sixty standout patents and it has been named as one of the top life changing discoveries arising from UK Universities (Universities UK)

Further reading

Chemiluminescence  reactions frequently involve the release of large amounts of energy. This energy is dissipated in the form of photons when electronically excited molecules, produced as a result of the reaction, relax to a stable ground state. The unique ability of chemiluminescence to provide a high sensitivity detection system has been exploited in medical diagnostics to provide tests with enhanced diagnostic accuracy and reliability. Today, one of the most widely used chemiluminescence detection techniques is Invitron’s Molecular Light Technology which employs the use of aryl Acridinium Ester (AE) compounds.

The technology for linking AE compounds to binding reagents such as antibodies or oligonucleotide probes is now well established. These labelled reagents can be designed to target specifically an infinite range of substances of biomedical interest which are then quantified by intensity of the light generated in the chemiluminescent reaction. Such assays were initially developed to investigate human diseases by analysing changes in blood constituents such as hormones. Subsequently, they have been applied to the detection and monitoring of tumour products, infectious agents such as HIV, markers of coronary heart disease and the gene mutations that give rise to inherited diseases. Acridinium based assays have been developed in a wide range of assay formats included fully automated systems (e.g. Centaur, Siemens Medical Systems; Tigris, Gen-Probe Inc) and microplate immunoassays. In the near future this technique will also be used to perform rapid diagnostic tests using portable instruments next to patients at the point of care.

These analytical methods are not confined to applications in the clinical laboratory. For example, that can also be used to assess the toxicological impact of environmental pollutants through monitoring changes in the expression of marker genes in organisms, tissues and cell lines.

Acridinium Ester labels are stable compounds which can be oxidised under alkaline conditions to generate intense chemiluminescence. They can be detected in a simple luminometer labels such as 125I. Light emission from acridinium esters is rapid, the reaction being completed typically in less than 1 second.

Since AE provides a direct labelling system, very few manipulations are required on enzyme labels, where it is necessary to convert substrates to detectable products following completion of the immunochemical reaction. With a measurement time of only 1 second, AE chemistry provides an extremely rapid and user-friendly detection system.

In ELISA systems, the working range of an assay is limited by the fact that light transmission decreases as an exponential rather than a direct function of product concentration, thus restricting the range over which measurements can be made. At low product concentrations, sensitivity is limited by the fact that transmission is maximal, making small changes in the signal difficult to differentiate. ELISA optimisation represents a trade between sensitivity and working ran photomultiplier tube can record light intensity in a linear fashion over at least 6 decades of product concentration against a background that is virtually zero. This means that assays based on over an extremely wide working range without compromising analytical sensitivity.   In comparison with fluorescence, where light energy itself is used to produce electronically excited states, the low background generated by chemilumin obtained with photon counting that makes chemiluminescence one of the most sensitive end systems.

The AE labels used in Invitron’s chemiluminescence detection system are stable compounds which provide high quantum light yields with extremely accurate quantification. Light emission at a wavelength of 425 nm occurs when the AE is oxidised in the presence of alkali to yield N-methyl acridone via a dioxetanone intermediate, according to the scheme shown in figure 1. A wide range of instrumentation is now commercially available for the sensitive measurement of chemiluminescence in both tube and microtitre plate assay formats. 

 In practice, the reaction is carried out by the use of hydrogen peroxide at alkaline pH. Since the peroxide is unstable at high pH, it is not practicable to prepare a single reagent to trigger the reaction. Two detection reagents are therefore used: • Reagent 1 is a dilute solution of hydrogen peroxide in 0.1 M nitric acid. • Reagent 2 is 0.25 M sodium hydroxide containing a surfactant.

 In a typical microtitre plate assay, reagents are incubated in antibody-coated wells. After washing to remove any unbound reagent, the following protocol is used to trigger the AE light emission in a luminometer:

 1. Add  Reagent 1 (100µl) 2. Time delay (approx. 2 seconds) 3. Add  Reagent 2 (100µl) 4. Measure light output using a luminometer (approx. 1 second) 

 This AE measurement procedure relies on the fact that mixing of the reagents takes place extremely rapidly and efficiently. Unlike other chemiluminescence based techniques that measure reaction rate, the entire flash of light produced by an AE light-off reaction is measured. This means precise and reproducible measurements can be made in just a few seconds.